E-ISSN 2548-0839
Volume : 2 Issue : 1 Year : 2024

Metrics

1.6
2023 IMPACT FACTOR
2.1
5 year Impact Factor
0.00061
Eigenfactor
3.4
2023 CiteScore
86/157
Journal Citation Reports (Clarivate, 2024)(Dentistry, Oral Surgery & Medicine (Science))
Quartile Q3
SCImago Journal & Country Rank
Comparison of the Cyclic Fatigue Resistance of Reciproc and Reciproc Blue Nickel-Titanium Instruments in Artificial Canals with Single and Double (S-shaped) Curvatures [Eur Endod J]
Eur Endod J. 2017; 2(1): 32 | DOI: 10.5152/eej.2017.17038

Comparison of the Cyclic Fatigue Resistance of Reciproc and Reciproc Blue Nickel-Titanium Instruments in Artificial Canals with Single and Double (S-shaped) Curvatures

Mehmet Adigüzel1, Berk Turgay2
1Department of Endodontics, Mustafa Kemal University School of Dentistry, Hatay, Turkey
2Department of Oral and Maxillofacial Surgery, Mustafa Kemal University School of Dentistry, Hatay, Turkey

Objective: The aim of this study was to compare the cyclic fatigue resistance of Reciproc (VDW, Munich, Germany) and Reciproc Blue (VDW) in artificial canals with both single and double (s-shaped) curvatures.
Methods: In total, 80 instruments were tested in this study. Reciproc R40 and Reciproc Blue R40 instruments were selected for the cyclic fatigue resistance test (n=20 in each test). Each instrument was rotated in canals with a single curvature (60° curvature, 5-mm radius) and a double curvature (first coronal curve: 60° curvature and 5-mm radius; second apical curve: 70° and 2-mm radius) until fracture. The time to fracture was calculated, and the length of each fractured fragment was recorded. A one-way analysis of variance (ANOVA) and Tukey’s post hoc tests were used to analyse the time to failure.
Results: All the instruments had significantly higher fatigue resistance in the single simulated curvature canal than in the double curvature canal (P<0.05). In both curvature groups, the time to fracture of the Reciproc Blue was longer than that of the Reciproc instruments (P<0.05). In the double curvature canal, the instruments tended to fracture more often in the apical curvature than in the coronal curvature. There was no significant difference in the length of the broken fragments between the two groups in either the single or double canal curvatures (P>0.05).
Conclusion: The Reciproc Blue instruments showed higher cyclic fatigue resistance than the Reciproc instruments in both single and double canal curvatures.

Keywords: Cyclic fatigue resistance, endodontics, instrumentation and rotary, reciproc, reciproc blue

Corresponding Author: Mehmet Adigüzel, Türkiye
Manuscript Language: English
×
APA
NLM
AMA
MLA
Chicago
Copied!
CITE
 (63 accesses)
 (944 downloaded)
LookUs & Online Makale